Khan, Shahed and Santos, Eleonora (2018): Dose Reference Levels for Pediatric Thorax CT: Can Image Quality Be Achieved While Maintaining Dose Reduction for Different Body Mass Indices? Published in: International Journal of Healthcare Education and Medical Informatics , Vol. 2, No. 5 (24 August 2018): pp. 10-13.
Preview |
PDF
MPRA_paper_88725.pdf Download (806kB) | Preview |
Abstract
Organ doses were calculated for diagnostic thorax CT, using data for 573 male Bangladeshi children from 0-10 years old, accounting for age and BMI. We have computed organ doses for an average male adult (170 cm*70 kg) undergoing a thorax CT examination using the ImPACT dose calculator and constructed ratios of organ dose to CTDIvol as a function of location of organs in the thorax, using a tube voltage of 120 kV for Siemens 64 slice CT scanner. BMIs are modelled as specific diameter water cylinders to obtain the factors for organ doses relative to the organ doses received by a reference adult. Results show that age operates a reduction in absorbed dose by a factor ranging from 0.33-0.79, which are consistent with other reference studies. Conversely, an increase of 1 BMI unit implies an increase in absorbed dose of up to 13% to maintain the quality standards of diagnostic imaging.
Item Type: | MPRA Paper |
---|---|
Original Title: | Dose Reference Levels for Pediatric Thorax CT: Can Image Quality Be Achieved While Maintaining Dose Reduction for Different Body Mass Indices? |
English Title: | Dose Reference Levels for Pediatric Thorax CT: Can Image Quality Be Achieved While Maintaining Dose Reduction for Different Body Mass Indices? |
Language: | English |
Keywords: | Dosimetry, Dose reference levels, Pediatrics, Thorax CT |
Subjects: | I - Health, Education, and Welfare > I1 - Health > I10 - General |
Item ID: | 88725 |
Depositing User: | Professor Eleonora Santos |
Date Deposited: | 07 Sep 2018 23:48 |
Last Modified: | 27 Sep 2019 14:53 |
References: | [1] DeMarco, J. J., Cagnon, C. H., Cody, D. D., Stevens, D. M., McCollough, C. H., O'Daniel, J., McNitt-Gray, M. F. (2005). A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms. Physics in medicine & biology, 50(17), 3989. [2] Huda, W., Scalzetti, E. M., Levin, G. (2000). Technique factors and image quality as functions of patient weight at abdominal CT. Radiology, 217(2), 430-435. [3] McCollough, C., Bakalyar, D. M., Bostani, M., Brady, S., Boedeker, K., Boone, J. M., McNitt-Gray, M. F. (2014). Use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT: The Report of AAPM Task Group 220. AAPM report. 6. [4] Menke J. (2005). Comparison of Different Body Size Parameters for Individual Dose Adaptation in Body CT of Adults.” Radiology. 236:565–571. [5] Turner, A. C. (2011). The Development of Methods to Assess Radiation Dose to Organs from Multidetector Computed Tomography Exams Based on Detailed Monte Carlo Dosimetry Simulations. University of California, Los Angeles. [6] Turner, AC, Zankl M, DeMarco JJ, et al. (2010). The Feasibility of a Scanner-Independent Technique to Estimate Organ Dose from MDCT Scans: Using CTDIvol to Account for Differences between Scanners. Med Phys. 37:1816–1825. [7] Zankl, M., Becker, J., Fill, U., Petoussi-Henss, N., & Eckerman, K. F. (2005). GSF male and female adult voxel models representing ICRP Reference Man–the present status. The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World, 1721, 1. [8] Zhou, H., & Boone, J. M. (2008). Monte Carlo evaluation of in infinitely long cylinders of water, polyethylene and PMMA with diameters from to. Medical physics, 35(6Part1), 2424-2431. |
URI: | https://mpra.ub.uni-muenchen.de/id/eprint/88725 |